Benchtop multiparameter PC 80 PRO. XS. Magnetic stirrer with arm. Probe: Hamilton Polylite Lab (pH), VPT 1 (cell), temperature probe. Cable AS7/BNC. pH and Cond solutions. Datalogger, software Datalink, USB keyboard. Measurement range: pH -2,00 to 20,00. COND 0,0 to 1000 mS (automatic). ISE 0.001 to 19999 ppm. TDS 0,1 mg/L to 100,0 g/L. Salinity 0,01 to 100,0 ppt. Temp -20 to 120 ºC. mV ± 2000 mV. Resistivity 1-10 mO.

XS
Packaging Reference Stock My Price
x u. XG-PC80P-3  Check availability Buy

Characteristic
Model : PC 80 PRO
Probe : Hamilton Polylite Lab VPT 1 conductivity cell in Platinum glass
Measurement range : pH -2.000 to 20.000 COND 0.0 to 1000 mS (automatic) ISE 0.001 to 19999 ppm TDS 0.1 mg/l to 100.0 g/l Salinity 0.01 to 100.0 ppt Temp. -20 to 120.0 ºC mV ± 2000 mV Resistivity 1-10 m?
Resolution : pH 0.1 / 0.01 COND 0.01 µS / automatic scale ISE- TDS - Salinity - Temp. 0.1 ºC mV 0.1 Resistivity -
Accuracy : pH ± 0.02 COND ± 1 % FS ISE TDS ± 1 % FS Salinity ± 1 % FS Temp. ± 0.2 ºC mV - Resistivity ± 1 % FS
Dimensions (mm) : 165x195x68
Weight (g) : 1050
Pack (u.) : 1

The Benchtop multiparameters allow all electrochemical measurements (pH, conductivity, TDS, temperature, etc.) to be carried out in a single device with less space and weight.

PC 8 PRO- Measurements: pH, redox, conductivity, TDS, temperature, salinity - GLP fuction: date, hour and calibration expiration memory recall of last pH and Cond calibration, date and time and informations about the condition of the sensors.- Supports DHS digital and analog sensors. Graphic display about the conditions of the electrode - Large colored display TFT 4/3'. Multiview. Simultaneous display of pH, conductivity and temperature- Automatic calibration pH 5 points and manual 2 points. Auto recognition of USA / NIST / DIN buffer solutions.- Automatic calibration ORP in one point (475 mV)- Automatic calibration COND 4 points Manual calibration on 1 point. Auto recognition of 84, 147, 1413 µS, 12,88, 111,8 mS conductivity standards- 2-ring conductivity cell. Selectable cell constant 0.1 / 1 / 10- Data logger 1000 GLP data points downloadable on PC with free software DataLink or printable- 2 USB ports for data PC downloading and for an external keyboard and RS232 port for printer connection- Numeric password for calibration management and data memory delete (1 user).- IP54 Rating- Includes magnetic stirrer with arm, sensor support, temperature probe, pH 7 and pH 4 buffer solutions, 1413 µS y 12,88mS conductivity standards and manual.- Power supply: AC/DC adaptor 12V / 1000mA

PC 80 PRO
- Measurements: pH, redox, ISE, conductivity, TDS, salinity, resistivity and temperature
- GLP fuction: date, hour and calibration expiration memory recall of last pH and Cond calibration, date and time and informations about the condition of the sensors.
- Supports DHS digital and analog sensors. Graphic display about the conditions of the electrode
- Large colored display TFT 4/3'. Multiview. Simultaneous display of up to 6 parametres at the user's choice
- Automatic and manual pH calibration up to 5 points. Auto recognition of USA / NIST / DIN / METTLER TOLEDO / MERCK / DIN/NIST buffer solutions.
- Automatic calibration with Ion (ISE) electrode, 5 points
- Automatic or manual ORP calibration in one point
- Automatic calibration COND 4 points Manual calibration on 1 point. Auto recognition of 84 , 147 , 1413 µS , 12,88 , 111,8 mS conductivity standards
- 2 and 4-ring conductivity cell. Selectable cell constant 0.1 / 1 / 10
- CAL TIMER: Calibration Timer: set frequency of calibration. Memory recall of last calibration, calibration alarm
- Compensation for Ultrapure water for low conductivity measurementd
- Data logger 8000 GLP data points downloadable on PC with free software DataLink or printable
- 2 USB ports for data PC downloading and for an external keyboard and RS232 port for printer connection
- User Management and Audit Trail, 4 access levels for 15 profiles, alphanumeric password (4 users)
- IP54 Rating
- Includes magnetic stirrer with arm, sensor support, temperature probe, pH 7 and pH 4 buffer solutions, 1413 µS y 12,88mS conductivity standards and manual.
- Power supply: AC/DC adaptor 12V / 1000mA